Faculty Operative Technique & Master Lecture: Safe Removal of Well-Fixed and Failed Acetabular Components in Revision Total Hip Arthroplasty (Douglas E. Padgett, MD).
1. High-Yield Summary
- Primary Clinical Indications: Explantation of an acetabular component is indicated for chronic periprosthetic joint infection (PJI) requiring two-stage exchange arthroplasty, progressive symptomatic aseptic loosening with circumferential osteolysis or migration, catastrophic polyethylene liner failure with metallic shell damage, recurrent prosthetic instability driven by severe component malorientation (>55 degrees inclination or excessive anteversion/retroversion), and adverse local tissue reactions (ALTR) associated with trunnionosis or modular interface corrosion.
- The Critical Move: Controlled tangential shearing of the bone-implant interface using specialized curved acetabular osteotomes or cylindrical Explant blades that strictly match the outer convex radius of the shell. Blades must hug the metallic implant contour throughout 360 degrees of rotation, initiating with short blades and progressing to full-depth blades to achieve complete circumferential release before applying any extraction force.
- The Catastrophic Technical Error: Applying high-torque cantilever or levering forces across an incompletely released, well-fixed porous component. This error propagates stress concentrations into structural pelvic columns, precipitating iatrogenic acetabular column fracture, acute pelvic discontinuity, or medial wall blowout with perilous intrapelvic instrument displacement into the external iliac or obturator vascular corridors.
- Host Bone Preservation Principle: Revision reconstruction durability depends directly on the quantity and structural viability of remaining host acetabular bone. Preservation of greater than 50% viable host bone contact permits stable biological fixation using porous hemispherical cups or modular porous augments, avoiding complex custom triflange cages or structural allografts.
2. Indications
Indications and Contraindications
Surgical decision-making regarding whether to retain or remove an acetabular component during revision total hip arthroplasty requires careful evaluation of component stability, position, locking mechanism integrity, and underlying pathology:
- Absolute Indications:
- Chronic periprosthetic joint infection (PJI) requiring complete implant and cement removal for biofilm eradication during two-stage exchange arthroplasty.
- Gross mechanical loosening evidenced by progressive radiographic migration, broken fixation screws, or complete circumferential radiolucent lines (>2 mm) across DeLee and Charnley zones I, II, and III.
- Recurrent prosthetic dislocation secondary to severe acetabular malposition (inclination >55 degrees, anteversion <10 degrees or >30 degrees) where modular elevated-rim or face-changing liners cannot re-establish a stable impingement-free arc of motion.
- Catastrophic mechanical failure, including complete polyethylene wear-through with severe metallic shell grooving, locking mechanism destruction, or shell fracture.
- Extensive retroacetabular osteolysis threatening structural column support beneath an unstable or compromised component.
- Relative Indications:
- Adverse local tissue reaction (ALTR) or metallosis from dual-modular neck taper corrosion or metal-on-metal articulation requiring component exchange to eliminate incompatible geometries or optimize orientation.
- Well-fixed cup in a patient with recurrent instability where exchange to a dual-mobility construct or constrained liner is planned, but the existing shell locking mechanism is obsolete, unidentifiable, or incompatible with contemporary cemented liner interfaces.
- Contraindications:
- A well-fixed, well-oriented cementless shell with an intact, undamaged locking mechanism in an aseptic revision setting where modular liner exchange alone accomplishes stability, bearing replacement, or osteolysis debridement.
- Medically unstable patient with profound systemic comorbidities unable to tolerate extended operative duration or anticipated blood loss.
- Active pelvic discontinuity where isolated component extraction without structural column stabilization and plating would induce catastrophic pelvic collapse.
3. Preoperative Planning
Imaging Review
Preoperative radiographic assessment requires a structured, multiplanar imaging protocol to define component fixation, orientation, remaining bone stock, and column integrity:
- Standard Radiographic Series: A low anteroposterior (AP) pelvis centered over the pubic symphysis, a true cross-table lateral of the affected hip, and 45-degree Judet iliac and obturator oblique views. The iliac oblique view profiles the posterior column and anterior acetabular rim; the obturator oblique view details the anterior column, posterior rim, and obturator foramen.
- Radiographic Signs of Looseness: Progression of radiolucent lines >2 mm in all three DeLee and Charnley zones, broken or bent transacetabular fixation screws, migration >2 mm superiorly or medially, change in abduction angle >5 degrees, or visible fracture of the surrounding cement mantle. Conversely, the absence of radiolucent lines and presence of stress-shielding osteopenia in Zone II with medial trabecular remodeling suggest rigid osseointegration.
- Advanced Metal-Artifact Reduction CT (MARS-CT): High-resolution CT imaging with 3D multiplanar reconstructions is essential in all revision acetabular cases. MARS-CT quantifies residual anterior and posterior column dimensions, detects retroacetabular cavitary defects hidden on plain radiographs, confirms medial wall breakthrough, tracks screw trajectories relative to retroperitoneal vessels, and rules out pelvic discontinuity by assessing cortical continuity across the sciatic notch and obturator ring.
Equipment Checklist
The operating room must be equipped with specialized explantation tooling and comprehensive revision reconstructive implants prior to incision:
| Instrument / Implant Category | Specific System / Tools Required | Intraoperative Function & Purpose |
|---|---|---|
| Curved Explantation Blades | Mechanical Explant System (Zimmer/Centerpulse) with full size-matched short and long cylindrical blades | Low-profile circumferential shearing of bone-implant interface hugging the outer metallic shell radius. |
| Manual Osteotomes & Gouges | Charnley and Moreland curved acetabular osteotomes, flexible thin osteotomes, 1/4-inch straight osteotomes | Controlled peripheral interface disruption, clearing bone overgrowth, and tangential cement mantle sectioning. |
| High-Speed Burrs | Midas Rex or pneumatic high-speed motor with carbide pencil burrs, egg-shaped burrs, and diamond burrs | Decorticating osteophytes, sectioning polyethylene liners, cutting stripped screws, and trimming cement. |
| Hardware Removal Set | Universal screw extraction system, reverse-helix conical extractors, trephines, square/hex drivers | Engagement and extraction of stripped, damaged, or deeply countersunk acetabular fixation screws. |
| Polyethylene Extraction Tools | Modular liner pullers, 3.5 mm self-tapping cortical screws with drill, threaded extraction pins, U-hooks | Disengaging locked polyethylene liners from the metallic shell without damaging host columns. |
| Revision Reconstruction Armamentarium | Highly porous titanium/trabecular metal revision shells, modular porous augments, revision screws, reconstruction cages | Immediate definitive reconstruction of anticipated Paprosky Type I through Type III acetabular bone defects. |
Examination Under Anesthesia (EUA)
Following induction of anesthesia and prior to sterile preparation, perform systematic manual examination of the affected extremity:
- Palpate pelvic stability and test for gross motion across the hemipelvis to identify unappreciated pelvic discontinuity.
- Record baseline passive hip range of motion, assessing fixed flexion, adduction, or rotational contractures that dictate extensive soft-tissue release.
- Verify baseline limb length discrepancies and soft-tissue tension across the abductor complex to guide subsequent reconstructive restoration.
4. Patient Positioning & Setup
Patient positioning must guarantee rigid pelvic stability while permitting full excursion of revision instruments and intraoperative fluoroscopy:
- Lateral Decubitus Position: Position the patient in the lateral decubitus position on a radiolucent operating table supported by a rigid pegboard or adjustable anterior and posterior pelvic positioners. Secure the anterior positioner firmly against the pubic symphysis and the posterior pad against the sacrum, avoiding lumbar lordosis or pelvic tilt that distorts visual orientation.
- Table Alignment Check: Ensure the long axis of the pelvis is strictly perpendicular to the floor. Subtle tilt into extension or flexion alters functional acetabular anteversion and inclination angles during component extraction and subsequent trial re-implantation.
- Peripheral Nerve and Pressure Padding: Place an axillary roll inferior to the dependent axilla to protect the brachial plexus. Pad the dependent peroneal nerve at the fibular neck and place protective gel padding between the lower extremities to prevent pressure necrosis.
Safety Check
Bring the C-arm fluoroscopy unit into the sterile field before final draping to confirm that unobstructed AP pelvis and Judet oblique views can be acquired without colliding with pelvic positioners or table attachments. Verify that high-volume suction, cell saver autotransfusion tubing, and rapid infusers are operational prior to skin incision.
5. Approach
The posterior approach (Kocher-Langenbeck / Moore) is the primary surgical workhorse for revision acetabular surgery, providing comprehensive circumferential exposure of the acetabular rim, posterior column, and retroacetabular space:
- Internervous and Intermuscular Boundaries: Dissection proceeds through the fascia lata and splits the gluteus maximus along its muscle fibers, which are innervated by the inferior gluteal nerve. The plane between the gluteus medius (superior gluteal nerve) and tensor fasciae latae is preserved superiorly. The piriformis, superior gemellus, obturator internus, and inferior gemellus are identified and reflected with stay sutures.
- Sciatic Nerve Mobilization and Protection: The sciatic nerve courses directly posterior to the short external rotators in the retrotrochanteric gutter. In revision procedures, extensive scar tissue frequently tethers the nerve directly to the posterior capsule or extruded cement. The sciatic nerve must be formally identified by palpation and direct visualization, mobilized from surrounding adhesions, and protected with a broad, dull malleable retractor or soft-tissue pad. The knee must be maintained in 30 to 45 degrees of flexion throughout the procedure to relax tension on the nerve trunk.
- Extended Exposure Corridors: If exposure of the acetabulum is obstructed by a retained femoral stem, perform an extended trochanteric osteotomy (ETO) or anterior capsular release to allow safe anterior displacement of the proximal femur without avulsing the abductor mechanism.
6. Technique
Step 1: Exposure and Soft-Tissue Mobilization
Achieving 360-degree visualization of the true acetabular rim is an absolute prerequisite before initiating component extraction:
- Incise the pseudocapsule circumferentially, resecting hypertrophic fibrous tissue and excessive scar envelope.
- Excise all retroacetabular osteophytes and overhanging bone lips using a rongeur or curved osteotome. Retained peripheral osteophytes mechanically entrap the shell rim, preventing axial dislodgement and forcing surgeons into dangerous levering motions.
- Position Retractors: Place an anterior Hohmann retractor over the anterior acetabular rim, maintaining bone contact to avoid femoral neurovascular injury. Position a superior Hohmann into the iliac wing above the dome, and place a bent Hohmann or Meyerding retractor inferiorly beneath the transverse acetabular ligament or into the obturator foramen.
Step 2: Liner and Hardware Removal
Component extraction must proceed in an orderly, stepwise fashion from bearing insert to underlying hardware:
- Extract the Polyethylene Liner: If the locking mechanism does not disengage with standard osteotome tapping, drill a 3.5 mm pilot hole in the non-articular dome of the insert. Drive a 4.5 mm or 6.5 mm cortical screw through the polyethylene until its tip contacts the metallic shell base. Advancing the screw acts as an internal jack, forcing the liner out of the locking rim. Alternatively, section the liner into quadrants with a high-speed pencil burr.
- Locate and Inspect All Screws: Clear fibrous tissue, bone ingrowth, and debris from screw head drive recesses using a fine curette or pick. Completely seat the appropriate screwdriver before applying torque.
- Back Out Supplemental Screws: Remove all transacetabular fixation screws. If a screw head is stripped, apply a reverse-threaded conical screw extractor with axial impaction, or trephine around the screw shank using a hollow core drill. Absolute Safety Check: Never attempt shell explantation until every screw has been confirmed completely removed; attempting extraction over a retained screw causes catastrophic column avulsion.
Step 3: Interface Disruption and Bone Preservation
The method of bone-implant interface disruption depends on whether the component is cementless or cemented:
- Well-Fixed Cementless Components (Explant System Technique):
- Select the cylindrical Explant blade whose curvature exactly matches the outer diameter of the shell.
- Anchor the central alignment mandrel into the apex dome hole or central driver thread of the shell to maintain concentric rotation.
- Engage the short blade first. Oscillate the blade back and forth while maintaining constant gentle inward pressure, keeping the cutting edge flush against the metal shell substrate. The blade shears trabecular osseointegration within 1 to 2 mm of the implant surface.
- Once the short blade reaches its maximum depth, swap to the longer curved blade to complete equatorial and polar release across the dome.
- Well-Fixed Cementless Components (Manual Curved Osteotome Technique):
- Utilize thin, flexible curved osteotomes (Charnley or Moreland). Orient the bevel of the osteotome toward the metallic shell and away from the host bone to direct cutting forces along the implant interface.
- Work circumferentially around the rim in 10- to 15-degree increments, advancing the osteotome only 5 to 10 mm per pass. Avoid driving osteotomes deep in a single quadrant, which risks cortical perforation.
- Cemented Acetabular Cups:
- Section the all-polyethylene cup into four quadrants using an osteotome or burr, then collapse and remove the segments.
- Create controlled radial stress-riser grooves in the cement mantle with narrow 5 mm osteotomes.
- Extract cement fragments tangentially using small gouges and curettes, fragmenting the mantle away from the underlying cancellous trabeculae without prying directly against the thin medial acetabular wall.
Step 4: Cup Extraction and Defect Assessment
Once circumferential interface release is complete, thread the revision cup inserter or vice grip securely onto the component rim or dome. Apply gentle rotational oscillations in internal and external rotation. The shell should rotate freely within the acetabular bed. Deliver the component axially without exerting cantilever levering against the peripheral rim.
Following extraction, thoroughly debride nonviable fibrous tissue, pseudomembrane, and particulate debris with curettes and pulsatile lavage to expose viable, bleeding host cancellous bone. Systematically evaluate remaining acetabular architecture to grade the defect according to the Paprosky Classification.
Critical Move
The single critical move is the continuous visual and tactile verification that the cutting blade remains strictly flush against the outer convex contour of the metallic shell throughout 360 degrees of rotation. The point of no return occurs during equatorial blade advancement: any outward deflection directs the blade away from the implant and deep into the structural bone of the anterior or posterior column, converting an easily manageable Paprosky Type I or II defect into an unstable Type III column loss or catastrophic pelvic discontinuity.
7. Pearls and Pitfalls
| Category | Technical Nuance / Strategy | Clinical & Biomechanical Rationale |
|---|---|---|
| Surgical Pearl | Pre-drill a 3.5 mm hole and drive a 4.5 mm cortical screw into the polyethylene dome against the metallic shell base. | Generates an internal jacking force that reliably pops stubborn, captured polyethylene liners out of intact locking mechanisms without rim damage. |
| Surgical Pearl | Maintain continuous sterile saline irrigation on Explant blades during motorized or manual oscillating passes. | Dissipates frictional thermal heat, preventing thermal osteonecrosis of adjacent host trabecular bone and extending blade sharpness. |
| Surgical Pearl | Advance the short Explant blade fully around 360 degrees before engaging the long blade. | Creates a concentric guide channel that prevents the longer blade from skiving obliquely into structural pelvic bone. |
| Surgical Pearl | Orient the bevel of manual curved osteotomes toward the metal implant rather than host bone. | Directs the cutting vector inward against the metallic surface, minimizing bone resection to a thin layer <1 mm. |
| Technical Pitfall | Applying heavy levering torque with a cup extractor when the component remains partially bound. | Induces massive posterior column fracture, converts stable walls into pelvic discontinuity, or causes catastrophic medial wall blowout. |
| Technical Pitfall | Attempting component extraction with an undetected or sheared screw shank still seated in bone. | Acts as an unyielding mechanical anchor that avulses a massive wedge of host bone when traction is applied. |
| Technical Pitfall | Using straight osteotomes tangentially across the acetabular rim to free deep polar ingrowth. | Straight blades cannot conform to spherical cup geometry and inevitably exit through the medial or posterior column cortex. |
| Technical Pitfall | Failing to identify and release peripheral retroacetabular osteophytes prior to shell extraction. | Creates a mechanical bottleneck that traps the equatorial diameter of the cup, resisting extraction even after complete interface release. |
8. Postoperative Care
Immediate Postoperative Protocol
Postoperative management is governed by the structural stability of the subsequent acetabular reconstruction:
- Weight-Bearing Status: If a rigid press-fit revision shell with robust multi-screw fixation was achieved in an intact acetabular bed (Paprosky Type I or II), patients may initiate weight-bearing as tolerated (WBAT) with a walker. If structural bone grafting, modular porous metal augments, or reconstruction cages were required for column loss (Paprosky Type III), restrict weight-bearing to flat-foot touch-down weight-bearing (TDWB, <20 lbs) for 6 to 12 weeks to protect biological osseointegration.
- Venous Thromboembolism (VTE) Prophylaxis: Administer dual-modality thromboprophylaxis combining intermittent pneumatic compression devices with direct oral anticoagulants (DOAC, e.g., apixaban 2.5 mg BID) or low-molecular-weight heparin (enoxaparin 40 mg daily) for 35 days postoperatively.
- Dislocation Precautions: Maintain standard posterior hip precautions (avoiding flexion >90 degrees, internal rotation, and adduction across the midline) for 6 to 12 weeks, especially in patients revised for recurrent instability or abductor compromise.
Rehabilitation Protocol
| Rehabilitation Phase | Timeline | Weight-Bearing & Activity Restrictions | Therapeutic Exercise Goals |
|---|---|---|---|
| Phase I: Protection | Weeks 0 to 6 | Touch-down weight-bearing (TDWB) or WBAT per reconstructive stability; walker/crutches | Ankle pumps, isometric quadriceps/gluteal sets, gentle active-assisted hip ROM within safety precautions. |
| Phase II: Transition | Weeks 6 to 12 | Advance to partial weight-bearing (50%), progressing to full weight-bearing with cane by week 10 | Closed-kinetic-chain strengthening, stationary cycling with high seat, abductor strengthening, balance exercises. |
| Phase III: Strengthening | Weeks 12 to 24 | Full weight-bearing without assistive devices; unconstrained community ambulation | Progressive resistive hip exercises, core stabilization, gait retraining, low-impact aerobic conditioning. |
| Phase IV: Maintenance | >24 Weeks | Full unrestricted daily activities; avoidance of high-impact ballistic sports | Long-term functional maintenance, joint protection education, annual clinical and radiographic follow-up. |
9. Outcome
Clinical and radiographic outcomes following acetabular component revision correlate directly with the amount of viable host bone preserved during explantation:
- Host Bone Preservation and Implant Survivorship: Preservation of greater than 50% viable host bone contact allows biological hemispherical press-fit fixation, yielding 10-year survivorship free from aseptic loosening exceeding 92% to 95%. When excessive host bone is sacrificed during explantation requiring structural allografts or massive cages, 10-year failure rates increase significantly, with mechanical failure reported in 15% to 25% of complex reconstructions.
- Functional Outcome Scores: Patients successfully revised with bone-preserving techniques achieve substantial gains in validated outcome metrics, with mean Harris Hip Scores improving from 45 to 52 preoperatively to 82 to 88 at mid-term follow-up. Pain relief is achieved in over 90% of patients without pelvic discontinuity.
- Comparative Explantation System Efficacy: Clinical studies evaluating mechanical curved blade explantation systems demonstrate significantly reduced operative time (mean reduction of 20 to 35 minutes) and significantly lower rates of structural column fractures compared to manual straight osteotome extraction techniques.
10. Complications
Acetabular explantation is associated with significant potential intraoperative and perioperative morbidities:
| Complication | Incidence Rate | Primary Mechanism & Risk Factors | Prevention & Rescue Strategy |
|---|---|---|---|
| Acetabular Column Fracture | 2.0% to 5.0% | Excessive cantilever levering against unreleased cups; eccentric osteotome advancement. | Use matched curved blades; identify resistance immediately; stabilize fracture with posterior column plates or modular augments. |
| Pelvic Discontinuity | 1.0% to 3.0% | Deep bone loss across the anterior and posterior columns connecting to the obturator ring. | Recognize discontinuity on CT; reconstruct with dual-column structural plating, porous distraction augments, or custom triflange cages. |
| Sciatic Nerve Neuropraxia | 1.5% to 3.5% | Excessive retractor traction, direct thermal injury from burrs, or entrapment in scar. | Formally mobilize nerve; maintain knee at 45 degrees flexion; avoid posterior retractor levering; release traction intermittently. |
| Major Vascular Laceration | <0.5% | Medial wall perforation or anterior retractor displacement into external iliac vessels. | Stay within safe Waschke zones; keep retractors directly on bone; avoid driving osteotomes through the medial quadrilateral plate. |
| Excessive Host Bone Loss | 4.0% to 8.0% | Straight osteotome skiving; aggressive high-speed reaming of well-fixed shells. | Utilize circumferential Explant system; shear bone within 1 mm of implant interface; assess defect before revision reaming. |
| Recurrent Instability / Dislocation | 5.0% to 10.0% | Abductor insufficiency, extensive capsular debridement, or revision component malorientation. | Restore native center of rotation and offset; verify 40-degree inclination and 20-degree anteversion; consider dual-mobility bearings. |
11. Anatomic Danger Zones & Pelvic Neurovascular Safety
Safe navigation of the acetabular perimeter during screw removal and blade explantation requires mastery of pelvic quadrant anatomy based on the Waschke coordinate system:
- Waschke Quadrant System: Constructed by drawing a primary line from the anterior superior iliac spine (ASIS) through the center of the acetabulum to the fovea, and a secondary perpendicular line bisecting the primary axis. This divides the socket into four distinct quadrants:
- Posterosuperior Quadrant (Safe Zone): Contains the thickest bone stock of the posterior column and iliac wing. Transacetabular screws can be placed safely up to 35 to 50 mm without exiting cortical bone. However, over-penetration beyond the sciatic notch endangers the superior gluteal neurovascular bundle.
- Posteroinferior Quadrant (Caution Zone): Contains the ischial bone bed. Screws placed here must not exceed 25 mm to prevent penetration into the lesser sciatic notch, which endangers the internal pudendal vessels and sciatic nerve.
- Anterosuperior Quadrant (Danger Zone): Directly abuts the external iliac artery and vein lying on the inner pelvic brim. Retractors placed over the anterior rim must never penetrate the iliopsoas belly. No fixation screws should ever be directed into this quadrant.
- Anteroinferior Quadrant (Danger Zone): Overlies the obturator canal containing the obturator nerve, artery, and vein. Screws directed here risk vascular laceration and catastrophic intrapelvic hemorrhage.
- Medial Wall (Quadrilateral Plate) Danger Zone: The medial acetabular wall is paper-thin (<2 to 3 mm) in dysplastic or osteolytic hips. Aggressive impaction of osteotomes or unguided burr debridement risks violating the quadrilateral plate, displacing instruments into the retroperitoneal space adjacent to the bladder and internal iliac venous plexus.
12. Acetabular Bone Loss & Defect Classification
Following successful explantation, acetabular bone deficiency must be classified to guide appropriate reconstructive strategy. The Paprosky Classification serves as the international standard:
| Paprosky Defect Type | Anatomic Characteristics & Rim Status | Ischial Lysis / Kohler’s Line | Recommended Reconstructive Strategy |
|---|---|---|---|
| Type I | Hemispherical rim fully intact; minor focal cavitary defects; columns rigid. | No ischial lysis; Kohler’s line intact; <1 cm migration. | Standard cementless hemispherical porous shell with multi-screw fixation. |
| Type IIA | Superior-medial rim distortion; intact columns; cavitary lysis <2 cm. | Mild ischial osteolysis; Kohler’s line intact; <2 cm superior migration. | High-hip center placement or standard porous shell with screw fixation and particulate bone grafting. |
| Type IIB | Superior-lateral rim deficiency <one-third circumference; columns intact. | No teardrop migration; Kohler’s line intact; <2 cm migration. | Jumbo hemispherical porous cup (≥62 mm) or modular porous metal buttress augment. |
| Type IIC | Medial wall blowout/deficiency; intact peripheral rim and columns. | Teardrop destruction; Kohler’s line breached; medial migration. | Porocoat/trabecular metal cup with medial impaction bone grafting or porous mesh. |
| Type IIIA | Severe superior rim loss (one-third to one-half circumference); posterior column intact. | Moderate teardrop lysis; ischial lysis <1.5 cm; 2 to 3 cm “up-and-out” migration. | Modular porous metal augments, porous revision shell, structural trabecular buttress. |
| Type IIIB | Massive column destruction (>one-half circumference); severe anterior and posterior column deficiency. | Severe ischial osteolysis (>1.5 cm); Kohler’s line disrupted; >3 cm “up-and-in” migration. | Cup-cage construct, custom triflange acetabular component (CTAC), or dual-column plating with augments. |
| Pelvic Discontinuity | Complete transverse fracture separating superior ilium from inferior ischiopubic segment. | Complete mobility of inferior hemipelvis; severe gap across posterior column. | Distraction porous augments with plating, double-column locking plate fixation, or custom triflange cage. |
13. Evidence-Based Synthesis
Modern clinical decision-making is grounded in high-quality randomized controlled trials and prospective registry analyses, corroborated by seminal biomechanical and anatomical descriptions.
| Clinical Question / Investigation | Level of Evidence | Pivotal Landmark Study | Study Design & Methodology | Cohort Size / Follow-up | Primary Outcome / Effect Size | DOI |
|---|---|---|---|---|---|---|
| What are the comparative mechanisms, efficacy, and bone-loss profiles of modern acetabular removal systems? | Level IV | Njoku I, Surace P, Piuzzi NS et al (2023) | Systematic Surgical Review | Comprehensive systematic review of modern mechanical explantation devices | Identifies curved blade removal systems as significantly reducing operative time and minimizing host bone loss compared to straight osteotomes. | DOI |
| How do bone preservation and complication rates compare between curved osteotomes and the Explant system during well-fixed cup removal? | Level III | Adelani MA, Goodman SB, Maloney WJ et al (2016) | Comparative Cohort Study | Comparative cohort of 68 patients undergoing well-fixed cementless acetabular component revision | Demonstrated that mechanical explant systems preserved host bone stock with lower rates of intraoperative acetabular column fracture. | DOI |
| What surgical techniques preserve acetabular bone stock during component removal in revision total hip arthroplasty? | Level IV | Paprosky WG, Weeden SH, Bowling JW Jr (2001) | Seminal Surgical Technique & Classification Series | 143 revision total hip arthroplasties evaluated for component removal and defect reconstruction | Established foundational classification of acetabular bone loss and safe explantation techniques avoiding pelvic discontinuity. | DOI |
| How effective is the Explant system in extracting monoblock or thin-walled well-fixed resurfacing cups without extensive pelvic osteotomy? | Level V | Bhutta MA, Gambhir A (2009) | Technical Innovation & Clinical Report | Technical series of challenging monoblock metal-on-metal cup explantations | Successful extraction of thin-walled well-fixed components with complete preservation of the native acetabular rim and columns. | DOI |
| What sequential surgical techniques optimize bone preservation during explantation of well-fixed cemented acetabular cups? | Level V | Stevens J, Macpherson G, Howie C (2018) | Surgical Technique & Biomechanical Review | Biomechanical and surgical analysis of controlled crack propagation and cement fragmentation | Controlled tangential cement fragmentation prevents deep column crack propagation and maintains stable structural bone bed. | DOI |
14. References
- Njoku I, Surace P, Piuzzi NS et al. Acetabular Component Removal Systems. JBJS Rev. 2023. doi:10.2106/JBJS.RVW.22.00246 [Level IV: Systematic Surgical Review]
- Adelani MA, Goodman SB, Maloney WJ et al. Removal of Well-Fixed Cementless Acetabular Components in Revision Total Hip Arthroplasty. Orthopedics. 2016. doi:10.3928/01477447-20160129-04 [Level III: Comparative Cohort Study]
- Paprosky WG, Weeden SH, Bowling JW Jr. Component removal in revision total hip arthroplasty. Clin Orthop Relat Res. 2001. doi:10.1097/00003086-200112000-00021 [Level IV: Seminal Surgical Technique & Classification Series]
- Bhutta MA, Gambhir A. Using the Explant Acetabular Cup Removal System for removing a well-fixed resurfacing cementless acetabular component. Ann R Coll Surg Engl. 2009. doi:10.1308/rcsann.2009.91.4.344a [Level V: Technical Innovation & Clinical Report]
- Stevens J, Macpherson G, Howie C. Bone preserving techniques for explanting the well-fixed cemented acetabular component. J Orthop. 2018. doi:10.1016/j.jor.2018.03.037 [Level V: Surgical Technique & Biomechanical Review]
15. Master Class Pro-Tip
When explanting a rigidly ingrown porous acetabular component, the master surgeon never uses the extraction handle as a lever against the peripheral bone rim. If resistance persists after completing circumferential blade passes, do not increase extraction torque. Stop and inspect the inner dome for hidden, sheared screw shanks or localized hydroxyapatite bridges anchoring the polar shell. Use a 2 mm pencil carbide burr to precisely disconnect stubborn osseous bridges at the bone-metal interface under direct vision. Preserving 2 millimeters of native subchondral bone stock converts what would be a complex custom triflange cage or structural allograft procedure into a straightforward biological jumbo cup reconstruction.
Master Class Surgical Pearl: Acetabular Component Explantation and Bone Preservation
Last Updated on September 23, 2026
🔗 Knowledge Graph & Clinical Correlates
OrthopaedicsOne Living Semantic Network🛠️ Operative Techniques
- Unicondylar Knee Arthroplasty Joint Reconstruction
🩺 Clinical Conditions & Indications
- Which EMR should I choose Original Clinical Contribution: Authored for the OrthopaedicsOne General Orthopaedics Specialty Section. Preserved, digitized, and updated to modern evidence-based...
🔬 Anatomy, Pathology & Diagnostics
- Why I Would Not Consider Navigating My THA Hip Joint & Proximal Femur
- Why I Would Consider Navigating My THA Hip Joint & Proximal Femur
- Why I Use Metal on Metal Hip Joint & Proximal Femur
- Why I Use Cross-Linked Polyethylene Hip Joint & Proximal Femur